Spec Graphs (4)

Created on: Wednesday, April 25, 2018
Graphs that display evidence that our specifications are met
  • Comparing different noise situations with a dummy cell using DETECT. Notice the current limits and the relatively small differences.
  • Comparison between DETECT and the industry standard on a dummy cell. Notice that in the regions of accuracy, the discrepancy between DETECT and the industry standard averages less than 1%
  • Comparison on DETECT of various chemical solutions. We have tested here distilled water, tap water, pickle juice, and our well known phosphate samples.
  • A long-experiment run on a dummy cell using DETECT. Note that the Normalized Mean Squared Error [ |(Data - Expect)/Expect|^2 ] is no larger than 0.035%. MEAN: 0.007557%, STDEV: 0.002384%